    Cylindrical Sterilizers
    Specifications

    Technical Specifications

    SAMBION 710 series — configurable across chamber sizes to match your CSSD throughput requirements.

    Inner Parts

    All inner parts are of Stainless Steel quality AISI 304/ 316/ 316L/ 316Ti or above (as per Customer’s Choice)

    Outer Body

    All outer body parts are SS and/or Powder Coated

    PLC Control Panel

    Fully automatic (MULTIPLE CYCLES) PLC control panel with online printer

    HMI System

    COLORED TOUCH-SCREEN HMI System

    Start Up Time

    Short start up time

    Cylindrical Sterilizers

    Operating Temperature

    Option to operate on 121 or 134°C (Any Temperature)
    Fitted with heavy duty water ring vacuum pump

    Air Expulsion

    Over 99.9% of air expulsion through purging

    Programmable Cycles

    8-20 programmable cycles option for varying load i.e.

    • Pre Start/ Warmup cycle
    • 2-6 cycles for different loads @ 121°C
    • 2-6 cycles for different loads @ 134°C
    • Leakage Test
    • Bowie+ Dick test
    • HTM/Helix Test
    • Customer’s Choice cycles @ 121°C
    • Customer’s Choice cycles @ 134°C

    Sliding/Hinged Door Mechanism

    Sliding/Hinged Door Mechanism (Double / Single) motorized

    Silicone Gasket

    High quality Silicone Gasket

    Door Sealing System

    Door sealing system (Pneumatic/Steam)

    HEPA Filter

    Fitted with HEPA filter

    Tray / Trolley Systems

    Stainless Steel Tray/ Trolley systems

    Boiler Water Feeding

    Easy Boiler Water Feeding System

    Cycle Completion Alarm

    With cycle completion Alarm

    Full Datasheet

    Major Specification Table

    Produced as model Sambion710 Series specifications

    Capacity (L) 75-150 L90-190 L115-290 L250-410 L470-570 L770-1880 L1150-2300 L75-5000L
    Chamber Size (mm) Dia 400 x L(600-1250)Dia 450 x L(600-1250)Dia 500 x L(600-1500)Dia 600 x L(900-1500)Dia 700 x L(1250-1500)Dia 900 x L(1250-3000)Dia 1000 x L(1500-3000)Custom Designed
    Chamber Material SS316L/304SS316L/304SS316L/304SS316L/304SS316L/304SS316L/304SS316L/304SS316L/304
    Sterilization Temperature 121-134°C121-134°C121-134°C121-134°C121-134°C121-134°C121-134°C121-134°C
    PLC + HMI Control
    Vacuum Drying
    Single / Double Door
    Door Type Sliding / Hinged
    Steam Generator Provided / Cauterized Boiler / both
    Thermal Printer Optional
    Automation Grade Automatic / Manual
    Phase Single / Three Phase
    Applications PLEASE ASK FOR THE CAPACITY & DUTY AND WE WILL OFFER MOST SUITABLE & ECONOMICAL OPTION
    CSSD/TSSU, Hospital, Pharmaceuticals, Laboratory & Bio-Medical Waste etc.
    Manufacture License CDSCO
    Standards BIS – IS:3829 Latest Edition with ISI mark
    Certificates IAF, NABCB, NSIC, MSME, ISO, CE (MDD), CE (PED), ISO 13485, ISO 9001 etc.
    ↓ Brochure Download Download Download Download Download Download Download Download

    Download Brochure

    Please fill in your details to download the brochure.

      Optional Features

      Add-Ons for Your Facility

      Automatic Boiler Water Feeding / Emptying System

      Steam Door Sealing System (instead of pneumatic)

      Online Circular Data Plotter (instead of online printer)

      Sterilized Air Injection for Vacuum Breaking Cycle

      Optional Models

      Alternative Configurations

      Beyond the standard fully automatic Class B machine, the SAMBION 710 series is also offered in the following configurations.

      Flash / Rapid / High Speed

      High-speed sterilization cycle option at 134°C for time-critical loads.

      Semi-Automatic

      Semi-Automatic

      121°C or 134°C operation with multi-port manual control, supplied without printer.

      Single Cycle Automatic

      Single Cycle Automatic

      Automatic machine running one fixed cycle at 121°C or 134°C, supplied without printer.

      Working Process

      A Validated, Five-Stage Sterilization Cycle

      The SAMBION 710 series uses a pre-vacuum cycle and/or gravity displacement cycle –
      air is actively drawn out of the chamber before steam is introduced, giving the steam  more direct and even contact with the load

      01

      Loading

      Trays/trolleys are racked and the motorized sliding door seals automatically.

      02

      Purging

      The water ring vacuum pump expels over 99% of air from the chamber.

      03

      Sterilization

      Saturated steam raises the chamber to the selected 121°C or 134°C cycle.

      04

      Exhaust

      Steam from the chamber is exhausted to atmosphere directly of via HEPA filters (as the case may be)

      05

      Vacuum Drying

      Vacuum-assisted drying removes residual moisture from the load. After complete drying the cycle completes and the load is ready for Unloading